2024.08.02 Recall: 700HPM High Pressure Calibration Manifold

06 August 2024

The following notice has been issued by the Office for Product Safety and Standards.

Product
700HPM High Pressure Calibration Manifold

Type
Pressure Equipment / Vessels - High Pressure Calibration Manifold

Brand
Fluke

Country of origin
United States

Product description
700HPM High Pressure Calibration Manifold.

Manifold

Product identifiers
The 700HPM (affected manifold) is a variable volume (VV) assembly designed to be used in closed-system pneumatic applications where changing the volume of the system will change the pressure. This affected manifold is supplied as a part / component to the Fluke 700HPPK (Kit) which is a manually operated high pressure pneumatic test pump. All serial numbers below 6516XXX are affected.

Hazard
This product presents a serious risk of injuries as during manual pressurization of the equipment, the plunger nut threads of the High-Pressure Pneumatic Calibration Manifold may yield releasing the plunger assembly and allowing it to be ejected at high velocity. The threads securing the plunger assembly have been found to fail prematurely. This can result in serious injury to the user.

The product does not meet the requirements of the Pressure Equipment (Safety) Regulations 2016.

Corrective action
Fluke UK Ltd has stopped shipping the affected manifold and recalled the product from end users. Further information is available on the Fluke website.